FAQ#110 - What's the purpose of the linear and polynomial factors written in the Calibration Reports?

Utilize the Linear Sensitivity Factors (A, B) and Ploynomial Sensitivity Factors (A, B, C, D) of the Calibration Reports permit to obtain readings in engineerig units with a maximum error as for the Calibration Report.

In the case of analogue gauges, the factors shall be applied on the electrical readings (i.e. mA, mV, digit, etc...)

In the case of digital gauges in "Legacy mode" (digital gauges with FW version 0.X, 1.X, 2.X or 3.X), the factors shall be applied on the output in engineering units (i.e. sin angle, kPa, etc...)

In the case of digital gauges in "Normal Mode" (digital gauges with FW version 3 or higher), the factors are already applied inside the gauge, so the Sensitivity Factors shall not be utilized with these instruments.
NOTE: all the new digital gauges are in "Normal mode".

For more information regarding SISGEO Calibration Report, refer to the following video https://youtu.be/BF8e3ZRcaYU

For more information on "how to use" the Sensitivity Factors with analogue gauges and with digital gauges in Legacy Mode, refer to the instruments manuals or to the following document: https://www.sisgeo.com/uploads/manuali/Addendum_data_processing/ADDENDUM_-_DATA_PROCESSING_-_EN_00.pdf

 

 

LT-Inclibus
MD-Profile system

 
FAQ#077 - Which are the maximum cable lengths from instrument to datalogger?

 

The cable lengths depend from many factors, first of all installation and cable protections (screen, earth, etc...). Assuming that instruments are installed in a workmanlike, for the maximum cable lengths should be followed the next suggestions:

DIGITAL INSTRUMENTS: see FAQ#073

ANALOGUE INSTRUMENTS: see both documents:

Cable length guide line with internal multiplexers
Cable length guide line with external multiplexers

 

 

 

LT-Inclibus
MD-Profile system

 
FAQ#029 - What is the Real Gauge Factor reported in the Specifications and Compliance Certificates of both VW embedded and arc-weldable strain-gauges?

The gauge factors for VW strain-gauges are generic factors calculated from the dimensions and the physical properties of the gauge material. The VW strain-gauges arep roduced in series with industrial method, so each strain gauge is equal to the other: it means that the gauge factor of VW strain gauges do not change.

The Gauge Factor is checked on some strain gauges from each manufacturing batch.

Vibrating Wire Strain Gauges

 
FAQ#028 - Are the vibrating wire strain-gauges and spot weldable VW strain gauges equipped with temperature sensors?

Yes, vibrating wire strain-gauges, as well as all other Sisgeo's vibrating wire instruments, are equipped with a built-in NTC thermistor YSI 44005 type (3000 Ohm at 25°C). Thermistor is included in the instrument price and therefore there is not additional cost for it.